QRS CARDS INTEGRATION PROCESS - 07.18.02 Initial integration of pcmcia cards demonstrated and found not to meet data collection requirements. Data collection requirements explained on this date allowing for better specification to programmer Setup, startup and module initialisation process shown graphically. ---------------------- Problem 1 OfficeMedic does not allow concurrent communication with EKG and PULSEOX cards. Solution NeatTools module has been created which talks to PULSEOX card, allowing collection of pulse rate and %sp02 directly from card concurrently with OfficeMedic performing EKG. Status OXI Data collection functionality demonstrated. ---------------------- Problem 2 Requirement for actual waveform from oximeter explained. Current implementation not reading value "SIGNAL INTENSITY" from oxicard. Solution NeatTools module guru developer informed of requirement Guru has oxi SDK for reference Status: Pending guru response ------------------------ Problem 3 Require data collection as close to sensor as possible ( ie raw ) to allow specific analysis to be performed by East Coast. Require sampling at 1024 Hz, or to be able to directly modify sampling of data Solution Neat Guru informed of requirements Status: Pending guru repsonse. Cross Reference with Problem 4 ---------------------------- Problem 4 Investigation of the QRS OfficeMedic software reveals that patient EKG data is stored into the MS-SQL database, with no immediate indication from the software that individual patient records can be extracted. Current protocol involves that complete patient database will be burned to patient unique CD-rom for dispatch via physical courier. Depending on actual physical size of test database (to be determined during testing protocol evaluation), this database may be uploaded to an ftp server (address to be determined) Solution (4A) Database sent to east coast analysis dept for investigation of data extraction potential. Possibilites include: - MS_SQL direct extraction of data ? Find someone who speaks SQL to answer this question. - Direct binary data extraction from MS-SQL ? - Contact with QRS manufacturer for information about data format and extraction from patient record Status (4A): PENDING (east coast evaluation) Solution (4B) NeatTools guru to be provided with EKG SDK. Request NeatTools module for concurrent EKG and OXI raw data collection Request sample rate as above to collect raw signals from cards, display within NeatTools, and store to non proprietary, accessible datafile Request timestamping of data points to allow precise correlation of different waveforms. Status (4B) : PENDING (west coast guru) --------------------------------------- Problem 5 Fingerprint measurement with Bioscrypt sensors : Assembly of components for interface with laptop. Solution Hardware integration expert to construct from supplied component parts. Status Parts dispatched to construction expert. Waiting. ----------------------------------------- Problem 6 Galvanic Skin Resistance measurement Solution NeatTools and TNG-4 hardware interface Status Awaiting delivery of TNG-4 hardware due Siggraph ----------------------------------------- RESOURCES: QRS OfficeMedic | PulseOxi SDK | EKG SDK VRMC registered as OEM developer with QRS, so assistance / technical query resolution may be possible directly from manufacturer or within SDK (assuming one reads and speaks SDK)